Bone-Marrow Edema Estimation with Microwave Imaging - Dataset

Study Reference

This dataset accompanies the study: "Bone-marrow oedema estimation with a microwave imaging OECP array in ex-vivo bovine long bones: comparison with CT ground truth"

Objective: Evaluation of a 16-element open-ended coaxial probe (OECP) microwave imaging array for non-ionising estimation of bone-marrow edema in ex-vivo bovine samples, validated against CT-based measurements.

Dataset Overview

The dataset contains microwave imaging (MWI), computed tomography (CT), and quantitative analysis results for seven ex-vivo bovine bone samples (A-G) from a single 13-month-old animal.

Samples:

  • 4 femur halves (distal and proximal from two femurs)
  • 3 humerus halves (from two humeri)

Experimental Design:

  • Simulated bone-marrow edema via precision-drilled holes filled with iodine contrast/gel mixture
  • Pre-injection and post-injection acquisition protocol
  • Ground truth validation through differential weighing and CT imaging

Sample Characteristics

Specimen Sample Type Drilled Holes Hole Diameter Hole Depth Nominal Volume per Hole
A Femur 16 2 mm 8 mm 25.13 µL
B Femur 16 2 mm 8 mm 25.13 µL
C Femur 25 2 mm 8 mm 25.13 µL
D Femur 25 2 mm 8 mm 25.13 µL
E Humerus 20 2 mm 8 mm 25.13 µL
F Humerus 20 2 mm 8 mm 25.13 µL
G Humerus 16 2 mm 8 mm 25.13 µL

Hole Pattern: 4 mm center-to-center spacing in a regular matrix pattern.

Acquisition Parameters

Microwave Imaging System

  • Array Configuration: 4×4 OECP array (16 probes)
  • Frequency Range: 2.25-3.00 GHz
  • Target Frequency: 2.5 GHz (analysis)
  • Nominal Power: 0 dBm
  • Calibration: OSL (Open-Short-Load) with air, aluminum foil, deionized water (22°C), and propylene glycol (22°C)
  • Spatial Grid: 16×16 effective measurements (array scanned over 16 positions with 4 rotations per position)
  • Acquisition Time: ~70 minutes per session (pre or post)
  • Raw Data: S₁₁ scattering parameters

Computed Tomography System

  • Scanner: GE Medical Systems Revolution Ascend
  • Parameters: 100 kVp, 80 mA
  • Reconstruction: BONE kernel
  • Slice Thickness: 3.0 mm
  • Pixel Size: 0.94 mm
  • CT Positivity Criterion: ≥50% voxels ≥250 HU

Directory Structure

GALEAZZI DATA/
├── summary-quantitative-results.csv
├── README.md
├── A/
│   ├── ct-scan-pre-injection.png
│   ├── ct-scan-post-injection.png
│   ├── dielectric-permittivity-matrix-pre-injection.csv
│   ├── dielectric-permittivity-matrix-post-injection.csv
│   ├── fluid-content-matrix-post-injection.csv
│   ├── dielectric-permittivity-image-A-pre-injection.png
│   ├── dielectric-permittivity-image-A-post-injection.png
│   ├── dielectric-permittivity-fusion-image-A-pre-injection.png
│   ├── dielectric-permittivity-fusion-image-A-pre-injection-with-values.png
│   ├── dielectric-permittivity-fusion-image-A-post-injection.png
│   ├── dielectric-permittivity-fusion-image-A-post-injection-with-values.png
│   └── roi-coordinates.json
├── B/
│   └── [same structure as A]
├── C/
│   └── [same structure as A]
├── D/
│   └── [same structure as A]
├── E/
│   └── [same structure as A]
├── F/
│   └── [same structure as A]
└── G/
    └── [same structure as A]

File Descriptions

Per-Case Folders (A-G)

Each specimen folder contains the following files:

CT Images

  • ct-scan-pre-injection.png - CT slice image before contrast injection (background removal applied)
  • ct-scan-post-injection.png - CT slice image after contrast injection (background removal applied)

Dielectric Permittivity Data Matrices

  • dielectric-permittivity-matrix-pre-injection.csv - 16×16 matrix of real relative permittivity values (εᵣ), pre-injection baseline
  • dielectric-permittivity-matrix-post-injection.csv - 16×16 matrix of real relative permittivity values (εᵣ), post-injection

Processing: Derived from S₁₁ parameters via Stuchly open-ended coaxial probe inversion model. Spatial resolution enhanced through nonlinear sigmoidal interpolation based on probe sensitivity characterization.

Fluid Content Matrix

  • fluid-content-matrix-post-injection.csv - 16×16 matrix of estimated aqueous fluid fraction (%)

Processing: Computed using CRIM (Complex Refractive Index Method) two-phase mixing model comparing post-injection permittivity against pre-injection baseline. Values weighted by probe penetration depth profile.

Dielectric Permittivity Images

  • dielectric-permittivity-image-{X}-pre-injection.png - Standalone visualization of pre-injection permittivity distribution
  • dielectric-permittivity-image-{X}-post-injection.png - Standalone visualization of post-injection permittivity distribution

Dielectric Permittivity Fusion Images

  • dielectric-permittivity-fusion-image-{X}-pre-injection.png - CT-MWI fusion image, pre-injection
  • dielectric-permittivity-fusion-image-{X}-pre-injection-with-values.png - CT-MWI fusion with dielectric permittivity values annotated at grid points
  • dielectric-permittivity-fusion-image-{X}-post-injection.png - CT-MWI fusion image, post-injection
  • dielectric-permittivity-fusion-image-{X}-post-injection-with-values.png - CT-MWI fusion with dielectric permittivity values annotated at grid points

Fusion Method: CT anatomical reference overlaid with MWI permittivity data for spatial co-registration.

Note: Images with "-with-values" suffix display the numerical dielectric permittivity values (εᵣ) from the corresponding permittivity matrix, mapped onto the fusion visualization at each grid point.

Region of Interest

  • roi-coordinates.json - Rectangular subset coordinates defining the analysis region for fluid volume computation

Root-Level Files

Quantitative Summary

  • summary-quantitative-results.csv - Aggregated volumetric measurements for all specimens

Columns:

  • Specimen - Sample identifier (A-G)
  • Sample_Type - Anatomical origin (Femur/Humerus)
  • Drilled_Holes - Number of contrast-injected holes
  • Measured_Injected_Volume_uL - Ground truth volume from differential weighing (µL)
  • Measured_Injected_Uncertainty_uL - Measurement uncertainty (µL)
  • Scaled_Injected_Volume_uL - Injected volume scaled to probe penetration depth (µL)
  • CT_Estimated_Volume_uL - Volume estimated from CT positivity criterion (µL)
  • CT_MWI_Scaled_Volume_uL - CT volume scaled to probe penetration depth (µL)
  • MWI_Measured_Volume_uL - Volume estimated from MWI fluid content analysis (µL)

Data Format Specifications

CSV Matrix Format

All CSV matrices are 16×16 grids representing the spatial distribution of measurements. Coordinate system:

  • Origin: Top-left
  • Row indexing: Top to bottom
  • Column indexing: Left to right
  • Missing values: nan (outside measurement region)
  • Units:
    • Permittivity matrices: Dimensionless (relative permittivity εᵣ)
    • Fluid content matrix: Percentage (%)

JSON Coordinate Format

ROI coordinates define rectangular analysis regions:

{
    "row_start": <int>,
    "col_start": <int>,
    "height": <int>,
    "width": <int>
}

Coordinates reference the 16×16 grid indexing system (0-indexed).