WatchPAT DATA Packet (0x0800) Payload Structure

Based on reverse-engineering actual device output with device worn on body

Last Updated: 2025-10-13


Overall Packet Structure

[24-byte Header: Standard WatchPAT packet header]
  - Magic: 0xBBBB
  - Command ID: 0x0800 (DATA packet)
  - Timestamp: Unix seconds
  - Transaction ID: Sequential
  - Length: Total packet size (header + payload)
  - CRC-16: Packet checksum

[Payload: 535-543 bytes (variable)]
  - 3-byte header
  - 6 AA AA delimited frames
  - Frame 5 contains 5 DD DD delimited samples

Payload Header (3 bytes)

Byte 0: 0x06 = Frame count (6 frames in packet)
Byte 1: 0x01 = Packet subtype/version
Byte 2: 0x00 = Reserved/flags

Assumption Confidence: HIGH - Byte 0 always matches the number of AA AA markers found


Frame Structure (AA AA Delimited)

Each frame starts with 0xAA 0xAA marker, followed by:

Byte 0: Frame ID (0x01-0x06)
Byte 1: Frame Type/Subtype
Bytes 2+: Frame-specific data

Frame 1, 2, 3: PPG Waveforms (Type 0x01/0x02/0x03 0x11)

Purpose: Photoplethysmography (PPG) waveforms from 3 LEDs for SpO2 calculation

AA AA 01 11 66 00 64 00 1E 02 00 00 [~96 bytes waveform data]
      ^^|^^ ^^^^^ ^^^^^ ^^^^^^^^^^^ ^^^^^^^^^^^^^^^^^^^^^^
      │ │  │     │     └─ Reserved (4 bytes)
      │ │  │     └─ Quality or Rate: 0x0064 = 100 (100%? 100Hz?)
      │ │  └─ Sample count: 0x0066 = 102 samples
      │ └─ Frame type: 0x11 = PPG waveform
      └─ Channel ID: 01=Red LED, 02=IR LED, 03=Ambient

Total: 112-120 bytes per frame (variable)

Frame Mapping

  • Frame 1 (0x01 0x11): Red LED waveform - for SpO2 calculation
  • Frame 2 (0x02 0x11): Infrared LED waveform - for SpO2 calculation
  • Frame 3 (0x03 0x11): Ambient light sensor - for noise cancellation

Waveform Data

~96-102 bytes of 12-bit PPG samples (device spec: 12-bit resolution)

Device Specification Reference

Per device spec:

  • Sample Resolution: 12 bits for PPG
  • Bandwidth: 0.1-10 Hz for PAT (PPG likely higher frequency)
  • Purpose: SpO2 requires multi-wavelength photoplethysmography (Red + IR)

Assumption Confidence: MEDIUM-HIGH

  • ✅ Three frames match multi-wavelength PPG requirement for SpO2
  • ✅ Sample count (~102) suggests ~1 second window @ 100Hz
  • ⚠️ Bytes 2-5 interpretation unclear (could be SpO2/HR or just metadata)

Frame 4: Status Frame (Type 0x05 0x10)

Purpose: Device status, battery level, LED currents, signal quality

AA AA 05 10 04 00 01 00 00 02 00 00 09 17 00 00
      ^^|^^ ^^^^^ ^^^^^ ^^^^^^^^^^^ ^^^^^
      │ │  │     │     └─ Status value: varies (e.g., 0x0280 = 640)
      │ │  │     └─ Flags: 0x0001
      │ │  └─ Data length: 0x0004 = 4 bytes
      │ └─ Subtype: 0x10
      └─ Frame type: 0x05

Total: 14 bytes (fixed)

Assumption Confidence: MEDIUM

  • ✅ Consistent structure across all packets
  • ⚠️ Exact meaning of status values unknown (battery? signal quality?)

Frame 5: Sample Container (Type 0x06 0x00) with DD DD Samples

Purpose: Container for 5 processed physiological samples

AA AA 06 00 50 00 05 00 00 00 00 00 [5 × DD DD samples]
      ^^|^^ ^^^^^ ^^ ^^^^^^^^^^^
      │ │  │     │  └─ Reserved (5 bytes, all 0x00)
      │ │  │     └─ Sample count: 0x05 = 5 samples
      │ │  └─ Data length: 0x0050 = 80 bytes
      │ └─ Subtype: 0x00
      └─ Frame type: 0x06

Total: 90 bytes (8-byte header + 82 bytes of samples)

Assumption Confidence: HIGH

Byte 4 (0x05) correctly predicts 5 DD DD markers in all observed packets


DD DD Samples (14 bytes each, 5 per packet)

Purpose: Processed physiological metrics derived from waveform analysis

CRITICAL OBSERVATION

All 5 samples share the same timestamp (e.g., 22435), indicating they are:

  • Multiple readings from the same time window
  • Quality-checked variations for reliability
  • Used to select best reading or calculate median

This is a quality assurance mechanism - the device runs its algorithm 5 times per second on the same data window and reports all results.

Sample Structure

DD DD A3 57 19 00 17 00 04 00 F4 FF 55 04 0F B3
      ^^^^^ ^^^^^ ^^^^^ ^^^^^ ^^^^^ ^^^^^ ^^^^^
      │     │     │     │     │     │     └─ [12-13] Checksum/Flags
      │     │     │     │     │     └─ [10-11] Sensor Reading (uint16)
      │     │     │     │     └─ [8-9] Accel Y or PAT delta (int16, signed)
      │     │     │     └─ [6-7] Accel X (int16, signed)
      │     │     └─ [4-5] Value2: Reserved (usually 0x00)
      │     └─ [2-3] Value1: Quality/Position/Stage
      └─ [0-1] Timestamp: Time window ID (uint16)

Field Interpretation (Based on Device-Worn Data)

Byte Field Type Observed Range Interpretation Confidence
0-1 Timestamp uint16 22435 (constant) Time window ID HIGH
2 Value1 uint8 16-33 Quality metric / Body position (5 states) / Sleep stage MEDIUM
3 Value2 uint8 0 Reserved HIGH
4-5 AccelX int16 -17 to +1 Actigraphy X-axis (12-bit, signed) MEDIUM-HIGH
6-7 AccelY int16 -12 to +3 Actigraphy Y-axis OR PAT delta (12-bit, signed) MEDIUM
8-9 Sensor uint16 1108-1110 PAT amplitude / Snore / Raw ADC value MEDIUM
10-11 Checksum uint16 Varies CRC or flags MEDIUM

Key Evidence from Device-Worn Data

1. Value1 (16-33 range)

Does NOT match SpO2 (70-100%) or HR (30-150 bpm)

Likely candidates:

  • Signal quality index (0-100 scale)
  • Body position (5 discrete states per spec: supine, prone, right, left, sit)
  • Sleep stage indicator (Wake, REM, Light, Deep)

2. AccelX/Y (negative values indicate movement)

Device spec confirms:

  • Actigraphy: 12-bit resolution
  • Chest Movement: 12-bit × 3 axis

Observed values:

  • AccelX: -17 to +1
  • AccelY: -12 to +3

Negative values = movement in specific direction

3. Sensor Reading (1108-1110)

Slowly incrementing across samples

Candidates:

  • PAT amplitude (device spec: 0.0-0.5V)
  • Cumulative counter
  • Raw ADC value from sensor
  • Snore intensity (12-bit per spec)

4. Five Samples Per Window

The device architecture appears to be:

  1. Collect 1 second of waveform data
  2. Run physiological algorithm 5 times with slight variations
  3. Report all 5 results
  4. Best/median result selected for clinical reporting
  5. Variance between samples indicates signal quality

Device Specification Cross-Reference

From device spec document:

  • Channels: PAT, Pulse rate, Oximetry, Actigraphy, Snoring, Body Position, Chest Movement
  • Sample Resolution:
    • PAT, Actigraphy, Snore: 12 bits
    • Oximetry: 1%
    • Body Position: 5 discrete states
    • Chest Movements: 12 bits × 3 axis

Assumption Confidence: MEDIUM

  • ✅ Field sizes and types validated
  • ✅ Signed values confirmed (negative movement)
  • ⚠️ Exact field meanings require testing with various activities (walking, position changes, etc.)
  • ⚠️ SpO2/HR location still unknown (possibly in PPG frame headers OR calculated separately)

Frame 6: PAT Waveform (Type 0x04 0x01)

Purpose: Peripheral Arterial Tone waveform for apnea detection

AA AA 04 01 46 00 64 00 00 00 00 00 [~70 bytes waveform]
      ^^|^^ ^^^^^ ^^^^^ ^^^^^^^^^^^ ^^^^^^^^^^^^^^
      │ │  │     │     └─ Reserved (8 bytes)
      │ │  │     └─ Quality: 0x0064 = 100
      │ │  └─ Sample count: 0x0046 = 70 samples
      │ └─ Subtype: 0x01
      └─ Frame type: 0x04

Total: 80 bytes (14-byte header + ~66 bytes waveform)

PAT Details

  • Bandwidth: 0.1-10 Hz (per device spec)
  • Sample Rate: ~70 samples suggests lower frequency than PPG
  • Purpose: Sleep apnea detection via arterial tone changes

Assumption Confidence: MEDIUM-HIGH

  • ✅ PAT is key metric for sleep apnea detection (device spec)
  • ✅ Lower sample count matches lower frequency requirement (0.1-10 Hz vs PPG's higher frequency)
  • ✅ Separate from PPG confirms multi-sensor approach

Unknown / Unverified Elements

1. SpO2 and Pulse Rate Location

Status: NOT FOUND in DD DD samples (values too low: 16-33)

Hypothesis A: Stored in PPG frame headers (bytes 2-5: 0x66 00 64 00)

  • Pro: PPG frames are the right place for SpO2-derived values
  • Con: Values (102, 100) don't match typical vital sign ranges

Hypothesis B: Calculated on-device but not transmitted in real-time

  • Pro: Android app doesn't parse DATA packets (saves raw)
  • Pro: Aligns with medical device architecture (edge sensing, cloud analysis)
  • Con: Real-time monitoring would require these values

Hypothesis C: Encoded differently (scaled, offset, or split across fields)

  • Example: SpO2 = (Value1 × 3) + offset
  • Example: HR in different frame or combined field

Needs: More device-worn data with varying vital signs (exercise, breath holding, etc.)

2. Value1 Exact Meaning (16-33 range)

Candidates:

  • Signal quality (0-100% scale, needs confirmation)
  • Body position (0-4 matching 5 discrete states)
  • Sleep stage (0-5: Wake, REM, Light, Deep, etc.)
  • Algorithm confidence score

Needs: Correlation analysis with:

  • Device position changes (lie down, sit up, turn)
  • Known sleep stages (if available from scoring software)

3. Sensor Reading (1108-1110)

Candidates:

  • PAT amplitude (0.0-0.5V per spec, scaled to uint16)
  • Snore intensity (12-bit value)
  • Raw ADC value from sensor
  • Respiratory movement amplitude

Needs: Analysis during:

  • Known snoring events
  • Apnea events
  • Various breathing patterns

4. Frame Size Variation

Observation: PPG frames vary 112-120 bytes

Hypothesis:

  • Waveform complexity affects encoding
  • Adaptive compression based on signal characteristics
  • Variable number of samples based on quality

Needs: Pattern analysis across longer recordings (hours)

5. Checksum Field (Bytes 10-11 in DD DD samples)

Status: Varies across samples (0x041B, 0x046F, 0x043A, etc.)

Hypothesis A: CRC-16 of sample data Hypothesis B: Flags and metadata packed into 16 bits Hypothesis C: Algorithm confidence or quality metric

Needs: CRC algorithm testing or pattern analysis


Data Flow Architecture

Device Sensors
    ↓
[PPG LEDs (Red/IR/Ambient)] → Frames 1-3 (102 samples @ ~100Hz)
[PAT Sensor]                → Frame 6 (70 samples @ ~10Hz)
[Actigraphy]                → DD DD samples (X/Y axes)
[Snore/Resp Sensor]         → DD DD samples
[Algorithm Processing]      → DD DD samples (5 readings/sec)
[Device Status Monitor]     → Frame 4
    ↓
BLE Transmission (0x0800 DATA packets @ ~1Hz)
    ↓
Android App (saves raw binary to file, no parsing)
    ↓
Upload to Server
    ↓
Server/Desktop Analysis Software (full parsing and scoring)

Architecture Confidence: HIGH

  • ✅ Matches Android app behavior (saves packets without parsing)
  • ✅ Aligns with medical device architecture (edge processing + cloud analysis)
  • ✅ Separates real-time data collection from offline analysis
  • ✅ Reduces mobile app complexity and battery usage

Example Packet Breakdown (Device Worn)

Packet Header

BB BB 08 00 DC 1E 00 00 00 00 00 00 0D 00 00 00 31 02 00 00 00 00 8B 32
│  │  │  │  └──────────────────┘  └──────────┘  └──────┘  │  │  └──────┘
│  │  │  │  Timestamp             TxnID: 13     Length:561 │  │  CRC
│  │  └──┘  CmdID: 0x0800                                  └──┘ Flags
└──┘ Magic

Payload (561 - 24 = 537 bytes)

06 01 00  [Packet header: 6 frames]

AA AA 01 11 66 00 64 00 1E 02 00 00 [...96 bytes PPG Red...]
AA AA 02 11 66 00 64 00 1E 02 00 00 [...96 bytes PPG IR...]
AA AA 03 11 68 00 64 00 12 7E 00 00 [...98 bytes PPG Ambient...]
AA AA 05 10 04 00 01 00 00 02 00 00 09 17 00 00
AA AA 06 00 50 00 05 00 00 00 00 00
  DD DD A3 57 18 00 15 00 04 00 F4 FF 55 04 0F B3
  DD DD A3 57 18 00 15 00 07 00 F2 FF 55 04 76 5A
  DD DD A3 57 12 00 18 00 07 00 F0 FF 56 04 E8 20
  DD DD A3 57 17 00 14 00 07 00 EF FF 56 04 66 3C
  DD DD A3 57 21 00 13 00 08 00 EF FF 55 04 0A B5
AA AA 04 01 46 00 64 00 00 00 00 00 [...66 bytes PAT waveform...]

Parsed Interpretation

  • 3 PPG Frames: 102 samples each @ ~100Hz (Red, IR, Ambient)
  • 1 Status Frame: Device metrics
  • 5 DD DD Samples: All timestamp 22435
    • Quality/Position: 24, 21, 24, 20, 33 (varying across samples)
    • Accel X: 4, 7, 7, 7, 8 (minimal movement)
    • Accel Y: -12, -14, -16, -17, -17 (consistent downward direction)
    • Sensor: 1109, 1109, 1110, 1110, 1109 (stable PAT?)
  • 1 PAT Frame: 70 samples @ ~10Hz

Summary of Confidence Levels

Component Confidence Reason
Overall structure HIGH Consistent across all packets
Frame delimiting (AA AA) HIGH Perfect match in all samples
Frame count in header HIGH Byte 0 always matches AA AA count
PPG frame purpose MEDIUM-HIGH Matches SpO2 requirements
DD DD sample structure MEDIUM-HIGH Field sizes validated
DD DD field meanings MEDIUM Plausible but unverified
Vital signs location LOW Not found in expected locations
Frame 4/6 interpretation MEDIUM Structure clear, semantics unclear
Data flow architecture HIGH Matches Android app behavior

Implementation Notes

The C# implementation (TelemetryDataStructures.cs, TelemetryHandler.cs) provides structured parsing based on these assumptions, with:

  1. Clear documentation of uncertainties
  2. Flexible field naming to accommodate multiple interpretations
  3. Complete raw data preservation for future analysis
  4. Event-driven architecture for easy testing and refinement

As more data becomes available (especially with varying physiological states), field interpretations can be refined without changing the parsing infrastructure.


Next Steps for Verification

High Priority

  1. Capture packets during position changes → Verify Value1 as body position
  2. Capture during exercise → Find SpO2/HR values (should be 120+ bpm, 95-100% SpO2)
  3. Capture during breath holding → See SpO2 drop below 90%

Medium Priority

  1. Capture during sleep → Correlate Value1 with sleep stages
  2. Capture during snoring → Identify snore intensity field
  3. Long recording analysis → Study frame size patterns

Low Priority

  1. Checksum verification → Test CRC algorithms on DD DD samples
  2. Compare with official analysis software → If available, compare parsed values
  3. Review firmware source → If accessible, verify field meanings

Document Version: 1.0 Date: 2025-10-13 Author: Reverse-engineered from WatchPAT ONE RE project Status: Working implementation with documented uncertainties