Calibration of CEM/AS3320 analog filter eurorack-modul.
Required or recommended tools are:
- 440Hz tone-reference (tuning-fork, fork oscillator etc.).
- Voltage meter.
- Frequency meter (if available).
- Eurorack for power (+12V/-12V) and keyboard-CV.
- 16pin Eurorack cable for modul-connection to the rack-bus.
- patchcables for modular synthesizers.
Table of Contents
Offset-adjustment is done with precision trimmer 'P_21' | 'P_22' on board B.
All following setups are done with Modul topside-potis.
- Set 'Frequency'-potentiometer to 12 o'clock position.
- Set 'Res.' (resonance)-potentiometer to value:10 ('CW', self oscillation).
- Set 'Mod' (modulation)-potentiometer to 12 o'clock position (0 := no modulation).
- Set all other potentiometers to CCW position.
- connect 16pin eurorack cable from rack-bus to the socket Pin1 on board B. Be carefull to fit the allignment of buspower Pin1 (-12V) .
- connect patchcable between signal-output 'OUT' and audio-out (or frequency meter, if available).
- power up the eurorack.
- Turn precision trimmer 'P_21' | 'P_22' on 1MoreF board B until the soundable- or measured- frequency is close as possible to 440Hz.
The CV-range (1V/octave) must be tuned with potentiometer 'P_11' | 'P_12' on board B after the offset-adjustment .
The 'Res.' (resonance)-potentiometer must be set to value:10 ('CW', self oscillation).
The following setups are done with Modul topside-potis and generating CV-voltages.
- Set the keyboard-CV for playing key-note A4.
- Set 'Track'-potentiometer to full CW.
- Turn 'Frequency'-potentiometer CCW until the soundable- or measured- frequency is again 440Hz.
- play key-note A3 and check the measured frequency. If the measurement is above 220Hz, then turn 'P_11' | 'P_12'-potentiometer 'CCW', else 'CW'.
- play key-note A5 and check the measured frequency. If the measurement is above 880Hz, then turn 'P_11' | 'P_12' -potentiometer 'CCW', else 'CW'.
Repeat this adjustment until the measuared frequencies are A2=110Hz, A3=220Hz, A4=440Hz and A5=880Hz.
Repeat also the 'offset-adjustment to fit the 440Hz to 12 o'clock position on 'Frequency'-potentiometer.
Voltage measurement on Testpoint: TP1.
The 'Res.' (resonance)-potentiometer must be set to value:10 ('CW', self oscillation).
Set all other potentiometers to CCW position.
- connect voltage-meter between GND and TP1.
The following table shows the measured voltages depending on the 'Frequency'-potentiometer position.
Poti-position | Frequency (Hz) | measured voltage on TP1 (mVolt) |
---|---|---|
min (CCW) | less then 20 Hz | 175 mVolt |
12 o'clock | 440 Hz | 84 mVolt |
max (CW) | above 20 kHz | -47 mVolt |
Hardware:cc by-nc-sa 4.0