From fa499394fdad953c06db550d5ee4093d27fccd4a Mon Sep 17 00:00:00 2001 From: BasedOnTechnology Date: Wed, 27 Oct 2021 11:18:56 -0400 Subject: [PATCH] Moved again --- .gitignore | 3 +- README.md | 637 ++++++++++++++++++++++++++++++++++++++++++++++++++++ generate.sh | 3 +- 3 files changed, 641 insertions(+), 2 deletions(-) create mode 100644 README.md diff --git a/.gitignore b/.gitignore index 0f9ee12..884ddfb 100644 --- a/.gitignore +++ b/.gitignore @@ -1 +1,2 @@ -BOM/ \ No newline at end of file +BOM/ +*.pdf \ No newline at end of file diff --git a/README.md b/README.md new file mode 100644 index 0000000..d31718d --- /dev/null +++ b/README.md @@ -0,0 +1,637 @@ +--- +bibliography: ../../documents/references.bib +--- + +Amplification-free fluorescent nucleic acid detection via synchronous photon counting +===================================================================================== + +![image](fluorescence/fluro_1){width="\\textwidth"} + +![Pardon the mess.](fluorescence/fluro_2){width="\\textwidth"} + +![image](fluorescence/light_source){width="\\textwidth"} + +![image](fluorescence/fiber_optic){width="\\textwidth"} + +![image](fluorescence/comparator){width="\\textwidth"} + +Quartus Prime 18.1 Lite edition + +Executive summary +----------------- + +Cuvette should be opaque or white to avoid autofluro; + +Initial state +------------- + +real-time + +After the dismal failure of luminescent infectivity quantification, and +the lack of success in infecting phage due to the small sample volumes +in use and the wrong phage type. The plaque assay took too long. + +No provision for magnetic shielding of the PMT was made. + +As with many other experiments in this project, many negative results +reported were tainted by the use of such a low sample volume. + +The typical method to detect. A quantitative PCR + +Such a device is known in biology as a plate reader. + +Nanodrop, using 280 nm absorbance. They're also \$10,000. + +Contrary to luminescence, you have control over when the excitation and +emission light turns on. This doesn't subtract effects like the +excitation light from filter leakage + +Conveniently, T4r has an extraordinarily large genome of approximately +172 kBp dsDNA(); each virion therefore For comparison, a fingerprint has +between 0.042 and 0.14 ng of DNA (). + +While these quantities are small, it is not particularly challenging, +and it is not our intention to suggest that this is a good design; we +are simply reporting on Designs for systems with comparable performance +are . The similar performance despite extremely high detector +sensitivity is probably due to the small light-collecting area due to +the objective, and + +Review +------ + +() quantifies adenovirus titer with a ssDNA 4.7 kbase genome. + +With a GelRed dye and 528/20 (note: BioTek filters are specified as +center wavelength / FWHM). + +() () offer excellent + +We show that both fluorescence and the excited state lifetime of SG +dramatically increase in viscous solvents, demonstrating an approximate +200-fold enhancement in 100 % glycerol, compared to water, which also +makes SG a prospective fluorescent viscosity probe. + +Biotium GelGreen has a very specific advantage: to increase the safety +of the dye, the flurophore is tied to some huge proprietary molecule, +preventing it from diffusing through membranes or capsids. This has the +side effect of making the fluorescence intensity strictly related to the +quantity of genomic material dispersed in the solvent, rather than + +Luckily, a recent paper has the answer: direct fluorescent detection of +DNA in solution, outside using dyes that bind to (intercalate into) DNA. +GelGreen doesn't penetrate. + +A similar method (using fluroescence microscopy rather than photon +counting) was also used by, and is generally a common practice in the +bioeffects field. + +Somewhat more challenging than viewing PCR output on a gel, since the +total quantity of DNA involved is quite low + +extra credit: how many photons are released? + +Unlike luminescence techniques, lock-in is possible + +Xu et al use + +Gel-Doc + +Flurophore +---------- + +the prototypical stain is Ethidium Bromide, but is challenging to obtain +outside certain laboratories. GelGreen is safe, very stable against +photobleaching and long-term storage, inexpensive, and readily +available. GelGreen is an Acridine orange (N-alkylacridinium) dye with a +similar spectra to green fluorescent protein. + +the base had integral overcurrent protection, which was triggered a few +times during development - a very useful + +For one thing, GelGreen appears to be eminiently stable - samples can be +stored for long periods of time, pre-mixed batches. + +bleaching was not obviously an issue. A calibration sample was stored in +a dark area with the dye bound to DNA for several months with less than +4% decay observed[^1]. + +tom Lexan cuvettes unexpectedly overwhelmed the DNA signal A surplus +Hammamatsu R4220 with HC123 current-limiting base at maximum sensitivity +was used. A low-voltage silicon photomultiplier like ON Semi's C-Series +SiPMs will probably be sufficient in most cases. + +(Phi6 uses an RNA - many dyes have different responses to +single-stranded (ss)DNA, dsDNA, or + +As noted by \[xi?\], this is a saturating effect; if too many +flurophores intercalate into the DNA the fluorescence is weaker. + +GelGreen is also sensitive to ssDNA and ssRNA but with 5 times lower +efficiency. GelGreen absorbs maximally at 272 nm and 507 nm and emits +maximally at 528 nm.() + +![Biotium GelRed/GelGreen fluorescence spectra. Credit Biotium Inc, +reproduced without permission.](gelred_gelgreen){width="50%"} + +Simple CMOS image stacking detection +------------------------------------ + +### () + +florescein An f1.2 lens is used. + +Notably, contrary to most arrangments, the excitation light was input +through the transparent bottom of the sample holders. They report that +\"the limiting factor is the \[filter leakage; in their paper they refer +to this as noise, distinct from image sensor noise\]\". + +An incidental advantage is that a color filter makes it easy to diagnose +issues with the light path. It is clear whether background, + +Lesson learned: color feedback + +a 30-s exposure on a cheap camera also almost got there. + +An ELP-brand camera USB100W05MT (a common choice in industrial systems) +with an OV9712 sensor was used. guvcview to capture. + +From luminescent techniques reported previously, DSLR sensors with long +exposure times are sensitive + +from \[\], using ImageJ () to stack. works great on cheap cmos cameras - +interestingly doesn't work at all on more expensive cameras. - almost +good enough - great for diagnostics + +interestingly, the fact that the camera has color is quite valuable; +filtering the green out can increase sensitivity a lot + +Photomultiplier photon wavelength discrimination +------------------------------------------------ + +Anyone familiar with photomultiplier tube use with + +lest you think I know what I'm talking about, I laboured under the +assumption that the pulse height somehwo + +() \"The photomultiplier tube outputs an electrical charge in proportion +to the amount of this scintllation, as a result, the output pulse height +from the photomultiplier \" + +\"Does anyone know of a circuit that can discriminate color PMT\"? + +only works if scintillator + +THE pulse height is equal to the input energy! The PMT can be made +color-sensitive; just subtract + +I thought the pulse height + +three counters? one above, one at? a lock-in amp would be better\... + +fast diode thresholding might work + +use one of the stm32f0 boards with comparators + +Possible artifacts and deficiencies in +-------------------------------------- + +This arrangement is patently unsuitable for producing scientific +results, as it has not even been calibrated against a DNA ladder. +Positive control samples were generated by cracking phage capsids of a +known titer in an autoclave and then mixing 1:1 with 1/4000 GelGreen. If +some other process + +The sample passed through a microfludic channel.[^2] Adsorption + +Instruments should not produce a continuous stream of results. + +Preparation and use +------------------- + +Undiluted GelGreen fluorophore (delivered at 10,000x concentration in a +neat little screw-cap) was kept at room temperature as advised (to +prevent crystallization or precipitation)[^3]. The fluorophore stock +solution was prepared by diluting 2.5 microliters of GelGreen in 10 mL +distilled water in a 15 mL screw-cap Falcon tube (McMaster-Carr +\#7979T33) producing a weakly orange solution of \"1/4000\" dilution and +stored at room temperature. Both were kept in light-tight metallized +bags when not in use. + +solution of GG in distilled water mixed 50/50 with the sample ("1/8000") +worked great in our case. + +The autosampler withdrew approximately 50 microliters of mixed solution +from the 0.4 mL stock tube (referred to as PG1), which was then injected +into an empty 1.5 mL empty sample tube. quantified. + +Cuvette +------- + +The custom 1 microliter slide cuvette used for initial testing was CNC +machined from 3 mm transparent Lexan-brand polycarbonate[^4]. After much +mystified head-scratching, it was found that the Lexan substrate was +very highly auto-fluorescent and completely overwhelmed the meagre DNA +signal. This occurred with a separate coupon of Lexan, but was not +observed with clear acrylic or polycarbonate. . That said, it has been +reported that polycarbonate does not auto-fluoresce significantly more +than acrylic (), so it is possible that some other effect led to this +result. + +1.5 mL Eppendorf-type clear polypropylene microcentrifuge tubes +(Carolina Premium Sterile Centrifuge tubes, 215245, believed to be MTC +Biotech SureSeal S). + +Microwell plates used for luminescence are usually opaque white to +reflect the few precious photons: fluorescence plates are typically +opaque black. + +Lesson learned: beware autofluorescence + +Light source +------------ + +Argon-gas lasers emit several closely-spaced lines in the visible +spectrum, the most prominent of which is at 488 nm. + +Cyan diode lasers emitting at 488 nm are SHARP GH04850B2G appears to be +obsolete. + +Due to their coherent emission, eye protection or suitable interlocks +are important when working with lasers. (Are tightly filtered LEDs +coherent?) + +lasers are available. + +() recommends + +### Diode laser tests + +Various impromptu tests were made with a 2.5W 445 nm laser cutter. +Commodity blue 445 nm emitters only barely clip the absorption spectrum +of GelGreen, apparently leading to a poor fluorescence yield. Combined +with the very high intensity, an unusably high background was found even +with the large 0.4 mL test sample. This observation were tainted by a +very poor optical arrangement and was largely inconclusive; 445 nm laser +diode alone was never tested with the final optical arrangement, and may +well have provided sufficient. This may allow the elimination of the +excitation filter. + +(Note that inexpensive laser cutter modules are often intensity +modulated via PWM rather than via analog current; this can cause great +confusion if not accounted for). + +gel doc papers use leds without filters, and paper says narrow leds +exist, talk about results with leds + +Blue LEDs \[Cree XLamp XP-E2 Blue Starboard\] are then sufficient for +excitation (though high CRI white LEDs emit more 480 nm blue, green +leakage is too high). + +Cree recommends staying below 300% of the continuous power when +modulating an LED. + +A white LED with a high color-rendering index (DK) seemed to have more +power in the blue passband; however, green leakage around the filter was +too strong. + +Arranging the light source physically at right angles can be +challenging. Using a plastic fiber optic assists in positioning +\"because of the small numerical aperture\". The fiber optic () + +Even a blue LED is visibly green through a + +Arranging the light source and detector optical paths at right angles is +the first line of defense against excitation light bleed-through. In +testing, is echoed by () + +> Finally, it is possible to excite a sample from one side and collect +> the fluorescence from the opposite side (Fig. 2.4.3D). While some +> instruments, such as microplate readers, use this approach, it is rare +> because it is difficult to completely prevent scattered excitation +> light from reaching the detector. +> +> Even when thin-film filters with extremely high blocking are used, +> high-angle scattered light can make it through the emission filter, +> since, as shown below, the spectrum of the filter shifts to shorter +> wavelengths for light at higher angles of incidence, thus causing the +> shifted emitter band to overlap with the excitation wavelength band. + +() \"These filters should also be specified to have very low ripple in +the passband, since the narrow laser lines of some lasers (especially +semiconductor diode lasers) can drift over time or with changing +environmental conditions, thus resulting in fluctuations of the filtered +laser power.\" + +Filters +------- + +Certainly the most critical aspect of any fluorescence technique is the +filter set used. + +Important characteristics include transmission % inside the passband, +optical density outside the passband, and sharp edges without long tails +crossing the so-called Stokes shift() between absorption and emission. +Some filters have ripple far from the edge of interest which must be +taken into account when assessing the overall filtering performance. + +() + +() contributes additional information + +Half-inch diameter laser line filters were used here for reasons of +cost. As of this writing, a set of quality GFP filters costs about 4 +times as much as a laser line set. Due to narrower pass-bands, perhaps +1/8 optical performance can be expected; in a photon-counting mode, +however, there appears to be ample sensitivity remaining. + +Superior filters can almost certainly be found; these were chosen for +convenience. + +Gel-docs , with spectra specified using the Wratten scale. These filters +alone do not seem to have optical density values sufficient for +amplification-free quantitation at these levels. + +Most commercial microscopes use dichroic beamsplitters, allowing the +excitation and emission beam to go through the same objective, a +technique known as epifluorescence (epi- means same-side). The dichroic +only removes the excitation to the 1% level - you still need the +dielectric filters, and they're really expensive. With fiber optic +excitation at right angles to the objective, excitation scattering was +low enough that the dichroic was unnecessary. + +The microscope itself was largely incidental, providing only a base. The +small focal length and aperture of the + +A 10/0.25 (10x magnification, 0.25 numerical aperture) objective was +used. + +Units reported as AU + +### Excitation filter used + +One filter, an + +(486 nm is the $n=4$ Balmer line for hydrogen). + +(note; this was a limited-stock clearance item that has since been +discontinued. Equivalent filters are readily available from other +suppliers, such as the ThorLabs FL05488-10 or Newport 05LF10-488). + +The anti-reflective side faced the LED; the reflective side faced the +fiber optic. + +ThorLabs FGB7 emission filter superficially looks satisfactory, but the +tails are believed to be too long to be useable. + +### Emission filters used + +Two filters in series, one: + +Note the confusion that can occur when specifying passband width as +$\pm$ versus FWHM. + +This is a long-pass edgepass filter. This is known as a Wratten or Gel +filter; the \#16 is known as the Wratten number. () has a table of +Wratten filter spectra; the cut-on wavelength (wavelength of 50% +transmission) is 530 nm for the \#15 and 540 nm for the \#16 ($\pm$ +circa 3 nm), with 90% maximum transmission. One layer of Kapton tape was +wrapped around the filter to protect other optics from the sharp edges. + +Tiffen-brand filters consist of two glass panes sandwiching a plastic +membrane that contains the dye proper. They can be cut to size + +The 10 nm FWHM emission filter is much narrower than professional GFP +filters, especially the super-wide edge-pass ones; a lot of photons are +lost that way, decreasing efficiency. However, it still appears to be +more than sufficient. + +manuals for gel-docs typically suggest \#16 or \#15. SYBR recommends +\#15. + +For an even lower-cost system, orange or amber acrylic sheets (typically +used for UV filtering) with similar filtering spectra (e.g. Acrylite +408-5) also exist; such a filter is used on the Carolina gel-doc, for +instance. + +This does mean is filtered before entering the fiber optic. + +in series with + +In general, gel filters appear to have a softer taper + +ThorLabs has series of colored glass (\"schott glass\") filters + +### Gel filter + +Unlike gels or plastic-glass sandwiches, thin-film filters have the +property that the pass-band depends greatly on the angle of incidence as +$$\lambda_{\text{shifted}} = \lambda \sqrt{1-\sin(\theta)^2}$$ + +The wavelength shifts shorter (bluer) as $\theta$ increases. This can be +a helpful property for creating tunable filters, but is a nuisance here. +This isn't an issue for the excitation filter, but is an issue for the. + +Thin-film dielectric filters also age; ThorLabs considers filters to +have a lifetime of 2 years. + +Since the LEDs and flurophore emissions are not naturally collimated, +this poses a little bit of an issue. The microscope objective seems to +provide sufficient collimation for the emission filter. + +The Edmund filter was unmounted glass. Also, these thin-film filters do +not extend to the edge of the glass. Edge-blackening (\"inked\") filters + +The wavelength can also be shifted by a few nm over the temperature +range 0 to 50 + +The LED was about 3 cm away, and put through a  3 mm aperture in a piece +of PVC pipe. The output from the filter was directly into the 1 mm +fiber, itself improving the bandpass. + +Electronics +----------- + +![image](fluorescence/PMT){width="\\textwidth"} + +Monitoring the output with a,[^5] a 100 ns switching time was typical. A +modulation frequency of 1 MHz was achievable, but no + +While drift in the power supply does not affect the background lock-in, +it can affect run-to-run repeatability. + +Running FPGA designs at varying speeds helps to debug race-condition +related bugs.[^6] + +() () + +take note of potentiometer positions + +sketches of light source in inkscape + +Atmospheric pressure glow discharge (APGD) plasma generation and surface +modification of aluminum and silicon si (100) + +huh! basic Argon discharge emits sharply at 488 nm - with a good fiter, +might work out! + +man, most white LEDs just have a notch taken out at 488! That's so +strange! + +The problem with lasers is that the only ones with power in the right +spectrum are from china or really expensive. + +Noncollimated light, the the band of a dielectric interference bandpass +filter shifts = $488 * (sqrt(1-(sin(45 deg)/2.08)^2)) = 458$; it only +shortens in wavelength. This would be a problem for the emission filter, +except we're using a colored-glass filter for that side. + +In fact, if we use a bunch of LEDs, this'll help to + +the CRI of the LED determines the amount of cyan light + +Clearance section at Edmund Optics is pretty great + +dichroic not needed - only does a first order-of-magnitude cut, mainly +important + +custom acrylic lightpipe? + +Time-domain or time-correlated photon counting +---------------------------------------------- + +An even better Phase-shift time domain fluorimetry. Iwata use a 20 MHz +DSO to measure a 5 ns $\tau$ fluorophore. However, this involves a light +source with a fast modulation bandwidth; and in the implementation they +describe, the PMT must be in the voltage mode. + +Called time-correlated photon counting. + +Another neat technique is to add I/Q + +Then your + +While there are ways to deconvolve a slow falling edge, there's another +problem: per time interval, the time spent in the excitation-off, +flurophore exponential decay time is proportional to the frequency of +the excitation light. If you're only getting 1000 photons from the +sample per second, the dye lifetime is 5 ns, and you're turning the +light on and off at 1 MHz, you're only getting a \[\] photons from the +exponential decay region. Even with a long exposure, that doesn't seem +to be enough to pick up the decay. + +It is also possible to perform flashbulb. () abuse of pmts. Note that as +long as average current limits are obeyed, PMTs are happy to endure very +high pulse currents, like flashbulbs for exciting; no shutters or +anything required. Gating the photomultiplier HV is another technique. +Figure out the average current limit for your PMT, the expected voltage +based on your load resistor value and watch that it never exceeds it. + +Another way to pick up this phase shift is to make your lock-in +amplifier phase-sensitive - very simple (a 2x clock put into a flip-flop +divider, then the middle). This also wasn't nearly good enough, +completely useless. + +quadrature input + +gel tampering + +() + +One might expect, given that white noise has an expectation value of +zero, that this technique would average out to 0, and the signal would +be linearly proportional to time. However, a 1-dimensional random walk +with a uniform step size of 1 is expected to end up $\sqrt{N}$ units +away from the origin after N steps. In the same time, a fluorophore +emitting at a rate of $r=1$ counts per step has an expected value of N +counts (a Poisson distribution has a mean equal to its parameter). +Therefore, $\text{SNR} = \frac{N}{\sqrt{N}} = \sqrt{N}$. Longer exposure +times would therefore provide better precision, but with diminishing +returns. + +A more comprehensive consideration () + +However, + +it has been noted that water is a good fluorescence quencher and that +this might be how + +If time-domain filtering is sufficient, + +Low background is required. + +Contrary to standard microscopy, you want as little excitation light to +enter the objective as possible. Using the existing + +the input edge is considered a clock; the minimum pulse width limits +apply, but are not clearly specified in the datasheet. In this case +(1/155 mhz) = 6 ns. + +However, we had no luck with precipitation. + +With the setup we're using and the small quantities, the excitation +light is somewhere around  $10^5$ times as powerful as the emission. +This doesn't seem to be a big issue gel-docs, picking out bands on gels +- they don't usually seem to use excitation filters; however, to get the +excitation bleed-through low enough to do this quantitative assay, the +bleed-through must be really low, and in our testing proper dielectric +filters are required on both the excitation and emission sides. + +There are a few sources of noise: + +PMT dark counts Some be filtered out by judicious use of comparator +pulse height threshold, the lock-in takes care of it. A non-issue in our +case. Ambient light leakage It'll be fine. Even with the room lights +Using microscope optics and the lock-in, essentially a non-issue for +use, surprisingly. Some fluorescence microscopes use micron-size +apertures to limit the depth of field to avoid Putting + +make Dia diagram of system + +The PMT was at maximum sensitivity (in fact, slightly above max voltage) + +You might be wondering why the filters are even required - why not just +subtract the excitation bleed-through with a control? Unfortunately, any +miniscule variation in the scattering of the excitation will be orders +of magnitude larger than the fluorescence signal you're looking for. + +Some papers discuss adding a third chopper or gate period or +photomultiplier or to measure the drift of the excitation light source. +Putting some feedback in the loop would probably + +Silicon photomultipliers like ON's C-Series are almost certainly +sufficient for this application, eliminating the HV requirements of PMTs +at the cost of a smaller active area, requiring a larger lens to collect + +half-life is 0.693 times the average lifetime. + +Simple spectrally-filtered intensity is good enough. + +Polarization is a neat way to filter; linearly polarize the excitation, +the emission comes out whatever orientation the DNA happens to be, which +is usually random. Apparently + +Performance and characteristics +------------------------------- + +Performance of this arrangement was very satisfactory. A 10-second +integration time, with, produced a background fluorescence signal of + 1500 counts, with per-sample stability of approximately $\pm 1500$ +counts. The + +Literature review +----------------- + +### () + +[^1]: pulse\_1.pnw line 2567 + +[^2]: pulse\_1.pnw lines 2517 + +[^3]: pulse\_1.pnw line 1367 + +[^4]: pulse\_1.pnw line 1719, 1755 + +[^5]: pulse\_1.pnw line 1879 + +[^6]: pulse\_1.pnw line 1879 diff --git a/generate.sh b/generate.sh index 1852e9a..e284997 100644 --- a/generate.sh +++ b/generate.sh @@ -1,2 +1,3 @@ #pandoc ../../documents/fluroescence.md -o fluorescence.html -pandoc --filter pandoc-citeproc --bibliography=../../documents/references.bib -s ../../documents/fluorescence.tex -o paper.md \ No newline at end of file +pandoc --filter pandoc-citeproc --bibliography=../../documents/references.bib -s ../../documents/fluorescence.tex -o README.md +pandoc --filter pandoc-citeproc --bibliography=../../documents/references.bib -s ../../documents/fluorescence.tex -o paper.pdf \ No newline at end of file